  Background

Despite the extreme high precision of the CTX biplanar fluoroscopy system, digitized data contains contain errors compounded from several potential sources. (1) The undistortion relies on accurately determining the centroids of the grid. Image quality and/or imperfection in flattening the grid may cause small error. (2) camera calibration provides a "best estimate" of camera position based on digitized clicks on the image and the "true" measured distances of the points on the calibration object. Digitizing error and/or offsets in the calibration object point distances can also add uncertaintly. (3) Digitizing, whether automated or by hand can also add error and finally (4) 3D reconstruction between the two cameras may also add error.

...